FlexiSpec® product line is the latest generation of Attenuated Total Reflection Near & Mid IR-fiber ATR-Probes produced for any type of FT-NIR, FT-IR and other IR-spectrometers, photometers and IR-LED or QCL spectral sensors. ATR immersion fiber optic probes with patented design are suitable for reaction monitoring in lab, pilot plant and for full automated process control.
Fiber Optic ATR-Probes for harsh environment are designed for process-spectroscopy in Near & Mid IR range to monitor reactions in-line in a broad temperature range from -100° to +250°C. They can resist to high pressures up to 200 Bar and used with FTIR or any other IR-spectrometers and spectral sensors in automated process control with process-interfaces.
Key Features
- High throughput in selected parts of Near & Mid InfraRed-spectra
- ATR-tip shaped for immersion in liquid flow without dead zone
- Robust for industrial applications in harsh environment up to 250°C
- Resistant to high pressure
- Compatible with all spectrometers and automated process-interfaces
Typical Applications
- Remote Reaction Monitoring in-line in temperature range -100/+250°C
- PAT applications in lab, pilot plant or industry with process-interfaces for automated process control
- Polymerization Process Control
- In-situ IR-Spectroscopy for PAT in Chemical, Petrochemical, Atomic, Bio-pharmaceutical & Food Industry
| Probe Type | Di-ART | Si-ATR | ZrO2-ATR |
| Transmission Range | 5.2-17 μm 600-1900 cm-1 | 5.2-17μm 600-3100 cm-1 | 1.5-6.5μm 600-3100 cm-1 |
| Fiber Type | PIR-900/1000 Silver Halide | PIR-900/1000 Silver Halide | CIR-500/550 Chalcogenide Glass |
| Temperature Range | -150°C / + 250°C | -150°C / + 250°C | -150°C / + 200°C |
| Max Pressure | 200 Bar | 100 Bar | 100 Bar |
Other Parameters
| Total Length | 1.5 m (1 to 5 m optional) |
| Shaft Length | 300 mm (100-500 mm optional) |
| Shaft Diameter | 12 mm |
| Shaft Material | Hastelloy C22 |
| Length of Legs | 500 mm (opt. 200 – 500 mm) |
| Protective Tube Material | Liquid Tight SS-Conduit, KOPEX-Tube |
| Minimal Bending Radius | 130 mm |
| Input/Output Connectors | Long SMA (opt.: any other type) |
| Cooling Air Flow Parameters | Excess pressure 0.5 Bar, Flow 2300 L/Hour |
| Inner Temperature Control | Inside shaft to control ART tip temperature |
Model Selection
| AP10344 | ATR-P-D-12-30-200/50-HT* Spectral Range: 5.2-17µm (600-1900cm-1) Fiber Type: PIR 900/1000 Tip Material: Diamond 200 Bar; -150°C/+250°C |
| AP10928 (Optional) | Tip Protection: crown sized – ATR-P-DC-12-30-200/50-HT |
| AP10932 | ATR-P-Si-12-30-200/50-HT* Spectral Range: 3.2-17µm (600-3100cm-1) Fiber Type: PIR 900/1000 Tip Material: Silicon 100 Bar; -150°C/+250°C |
| AP10512 (Optional) | Tip Protection: crown sized – ATR-P-SiC-12-30-200/50-HT |
| AP10935 | ATR-C-ZrO2-12-30-200/50-HT Spectral Range: 1.1-6.5µm (1550-9000 cm-1) Fiber Type: CIR 500/550 Tip Material: Cubic Zirconia 100 Bar; -150°C/+200°C |
| AP10936 (Optional) | Tip Protection: crown sized – ATR-C-ZrO2C-12-30-200/50-HT |
| Thumb | File information |
| Overview-ATR-Probes.pdf File Size 308.53KB | |
| ATR-Probe for Harsh Enviroment.pdf File Size 1.12MB |
Art Photonics, FlexiSpec® Fiber Optic Probes, High Temperature for Harsh Environment
- Brand: Art Photonics
- Product Code: AP10344
- Availability: Pre-Order
Related Products
Art Photonics, FlexiSpec® Raman Fiber Probes, Multi-wavelength Excitation (630-785 nm) and Single-wavelength Excitation (532 and 785 nm)
FlexiSpec® product line includes the high sensitivity Raman fiber optic probe to be used with any Ra..
Art Photonics, FlexiSpec® NIRaman Combi Probes, Multi-channel Fiber Optic Probes
FlexiSpec® product line includes innovating NIRaman Combi probes. Raman and NIR spectra contai..
Art Photonics, FlexiSpec® Reflection Fiber Optic Probes, (7+1) Design
Reflection Fiber Optic Probes from FlexiSpec® product line are designed to measure Diffuse and Specu..
Art Photonics, FlexiSpec® Transflexion Dual Pass IR Fiber Optic Probes for Gases
FlexiSpec® product line includes the latest generation of Transflexion (dual pass) fiber optic probe..
Art Photonics, FlexiSpec® Sterilizable ART Probes, Shaft-in-Shaft Design(12.0 mm OD)
Sterilizable ATR Fiber Probe is a new product of FlexiSpec® product line. These ATR Probes are produ..
Art Photonics, Flat Tip ATR Touch Probe, Flat Tip Diamond (600-1900 cm-1) or Germanium (600-3100 cm-1)
The flat tip ATR crystal design allows touch measurement of sensitive samples and soft solids like s..
Tags: Spectrometer, Microscopy, FTIR, Raman, Sample Preparation, Photonics, Fiber Probe