• Acoustic and Electro-acoustic Spectrometer DT 1202
    Acoustic and Electro-acoustic Spectrometer DT 1202 Acoustic and Electro-acoustic Spectrometer DT 1202 Acoustic and Electro-acoustic Spectrometer DT 1202

DT-1202 is an instrument designed for performing particle size distribution and zeta potential to characterize concentrated dispersions, emulsions and liquids. The measuring chamber of DT-1202 contains the both acoustic and electro-acoustic sensors. As a result this instrument is capable of measuring an attenuation spectra for for particle size as well as Colloid Vibration Current for zeta potential. Also, this instrument can perform acoustic and electro-acoustic measurements separately and independently.

Key Features 

  • Particle size distribution
  • Zeta Potential and Iso-electric point
  • Porosity and solids content
  • Compressibility and bulk viscosity
  • Non-aqueous conductivity and debye length
  • Acoustic attenuation and sound speed




The DT-1202 can be supplemented with a set of Options  which substantially extend this instrument's capabilities.

More information can be found in the Specification tab.







Calculated Parameters

Mean size [microns]
0.005-1000
Zeta Potential [mV]
+/- (0.5%+0.1)
Weight Fraction/Porosity
+/-0.1%
Compressibility E10 [1/Pa]
+/-0.003
Bulk Viscosity [cP]
+/-0.01
Debye length [nm]
+/-0.1
Measured Parameters

Temperature [Deg C]
0 to 100 , +/- 0.1
pH
0.5-13.5, +/-0.1
Frequency Range [MHz]
1-100
Ultrasound Attenuation [dB/cm/MHz]
0 to 20, +/- 0.01
Sound Speed [m/sec]
500-3000, +/-0.1
Ellectroacoustic signal [mV(s/g)1/2]
+/-1%
Conductivity [S/m]
10^(-11) to 1, +/- 1%
Measurement Time [min]
0.5-10
Sample Volume, Minimum [ml]

Size only
15
Zeta only
0.1
Both, no sedimentation
15+.01
Both with mixing
70
Both with titration
110
Both with pumping
150
Sample Requirements

Volume fraction, % (1)
0.1-50
Conductivity
N/A
pH
0.5-13
Temperature [Deg C]
<50
Viscosity of media [cP] (2)
<20,000
Viscosity of sample
<20,000
Particle size [microns]
0.005-1000
Zeta potential [mV]
N/A
Physical Specifications

Electronic unit weight [Kg]
20
Sensor unit weight [Kg]
30
Voltage [V]
100-250 VAC, 50-60 Hz
Power [W]
<300
Software
Embedded Windows HP, MS Office
More information can be found in the Datasheet.

(1) Instrument can measure attenuation well above 50% vl, but verification of the theory for computing particle size and zeta potential is not possible above this limit.
(2) The “micro-viscosity” is important for theoretical calculation. It might be different than “macroscopic” viscosity for gels and other structured systems measured with conventional rheometers.